Gas Data GM25-3 Gas Analyser
Landfill Gas & Extraction System Analyser
The Gas Data GM25-3 is a portable landfill gas analyser designed for simultaneous measurement of gas composition, pressure and calculated gas flow.
The fixed configuration measures methane (CH₄), carbon dioxide (CO₂), oxygen (O₂), hydrogen sulphide (H₂S) and carbon monoxide (CO), together with atmospheric, static and differential pressure. Gas flow can be calculated from differential pressure across an orifice plate, providing flow measurements in Nm³/hr.
This makes the GM25-3 particularly suited to landfill gas extraction and control systems, where gas composition, pressure and higher extraction flow rates need to be assessed using a single field instrument.
Key Features
Five gas channels: CH₄, CO₂, O₂, H₂S and CO
Atmospheric pressure measurement
Static pressure measurement
Differential pressure measurement
Calculated flow measurement using differential pressure
Ideal for landfill gas extraction systems
ATEX, IECEx and UKEx rated
Internal data logging
DataField GM25 Connect compatibility
Optional gas velocity and temperature measurement
8–12 hour battery life
12-month service interval
Application guidance: Choose the GM25-3 primarily for landfill gas extraction wells and gas collection systems, particularly where flow through an extraction system needs to be calculated from differential pressure.
Specification | Details |
|---|---|
ATEX Rating | ATEX, IECEx, UKEx II 2 G Ex ib IIB T1 Gb |
Dimensions | 260 × 100 × 75 mm |
Weight | 1.1 kg |
Ambient Temperature Range | -10°C to +40°C |
Aspiration Rating (minimum) | 300 mL/min |
Gas Measurements
Channel | Sensor Type | Range | Resolution | Accuracy |
|---|---|---|---|---|
CH₄ | IR | 0–100% | 0.1% | ±3.0% vol. (0–100%); ±0.3% vol. (0–5%) |
CO₂ | IR | 0–100% | 0.1% | ±3.0% vol. (0–100%); ±1.0% vol. (25–45%) |
O₂ | EC | 0–25% | 0.1% | ±0.5% vol. |
H₂S | EC | 0–5,000 ppm | 1 ppm | ±5.0% of reading or 75 ppm, whichever is greater |
CO | EC | 0–2,000 ppm | 1 ppm | ±5.0% of reading or 75 ppm, whichever is greater |
Pressure & Flow Measurements
Measurement | Sensor Type | Range | Resolution | Accuracy |
|---|---|---|---|---|
Atmospheric Pressure | Transducer | 800–1,200 mbar | 1 mbar | ±5 mbar |
Static Pressure | Transducer | ±250 mbar | 0.1 mbar | ±2.5 mbar |
Differential Pressure | Transducer | ±1,250 Pa | 1 Pa | ±3 Pa to ±250 Pa across range |
Flow Rate | Calculated from differential pressure | Application dependent | — | Application dependent |
The GM25-3 calculates flow from the differential pressure measured either side of an orifice plate, with flow displayed in Nm³/hr and is therefore best suited for higher flow extraction systems
Optional Measurements
Measurement | Sensor Type | Range | Resolution | Accuracy |
|---|---|---|---|---|
Velocity | Vane anemometer | 0.7–40 m/s | 0.1 m/s | ±0.5 m/s |
Temperature | Thermistor | -10 to +100°C | 0.5°C | ±1.0°C |