Gas Data GM25-2 Gas Analyser
Configurable Portable Gas Analyser
The Gas Data GM25-2 is a versatile portable gas analyser that can be configured to suit a wide range of environmental and process gas monitoring applications.
Unlike the fixed-configuration models in the GM25 range, the GM25-2 allows the required gas channels to be selected for the application. Infrared sensors are available for methane (CH₄) and carbon dioxide (CO₂), while electrochemical sensors can be configured for oxygen (O₂), carbon monoxide (CO), hydrogen sulphide (H₂S) and hydrogen (H₂). Optional pressure measurement can also be incorporated.
This flexibility makes the GM25-2 particularly suitable where the standard landfill gas configuration is not appropriate, including biogas, wastewater treatment and odour monitoring applications.
Key Features
Configurable gas analyser
CH₄ and CO₂ infrared measurement options
O₂, CO, H₂S and H₂ electrochemical sensor options
Up to six measured gas channels
Optional LEL and N₂ calculations
Optional atmospheric, static and differential pressure measurement
ATEX, IECEx and UKEx rated
Internal data logging
DataField GM25 Connect compatibility
8–12 hour battery life
User-replaceable battery pack
12-month service interval
Application guidance: Choose the GM25-2 where you require a custom gas configuration, particularly for biogas plants, wastewater and sewage treatment, or specialist gas monitoring.
Specification | Details |
|---|---|
ATEX Rating | ATEX, IECEx, UKEx II 2 G Ex ib IIB T1 Gb |
Dimensions | 260 × 100 × 75 mm |
Weight | 1.1 kg |
Ambient Temperature Range | -10°C to +40°C |
Aspiration Rating (minimum) | 300 mL/min |
Gas Measurements – Configurable Options
Channel | Sensor Type | Range | Resolution | Accuracy |
|---|---|---|---|---|
CH₄ | IR | 0–100% | 0.1% | ±3.0% vol. (0–100%); ±0.3% vol. (0–5%) |
CO₂ | IR | 0–100% | 0.1% | ±3.0% vol. (0–100%); ±1.0% vol. (25–45%) |
O₂ | EC | 0–25% | 0.1% | ±0.5% vol. |
H₂S | EC | 0–5,000 ppm | 1 ppm | ±5.0% of reading or 75 ppm, whichever is greater |
CO | EC | 0–2,000 ppm | 1 ppm | ±5.0% of reading or 75 ppm, whichever is greater |
H₂ | EC | 0–1,000 ppm | 1 ppm | ±5.0% of reading or 50 ppm, whichever is greater |
LEL | Calculated | 0–100% | 0.1% | Indication only |
N₂ Balance | Calculated | 0–100% | 0.1% | — |
Optional Pressure Measurements
Measurement | Sensor Type | Range | Resolution | Accuracy |
|---|---|---|---|---|
Atmospheric Pressure | Transducer | 800–1,200 mbar | 1 mbar | ±5 mbar |
Static Pressure | Transducer | ±250 mbar | 0.1 mbar | ±2.5 mbar |
Differential Pressure | Transducer | ±1,250 Pa | 1 Pa | ±3 Pa to ±250 Pa across range |
Optional Measurements
Measurement | Sensor Type | Range | Resolution | Accuracy |
|---|---|---|---|---|
Velocity | Vane anemometer | 0.7–40 m/s | 0.1 m/s | ±0.5 m/s |
Temperature | Thermistor | -10 to +100°C | 0.5°C | ±1.0°C |