Soil Redox Probe
The ecoTech Redox Electrode according to Mansfeldt is designed for continuous measurement of redox potential (Eh/ORP) directly within soils. Suitable for permanent field and laboratory installations, the electrode can operate under both unsaturated and water-saturated conditions.
The compact probe incorporates a normalised 99.95% platinum measuring surface, designed to provide good contact with minimally disturbed soil and improve reproducibility between measurement locations. The measuring electrode is separated from the reference electrode, allowing the reference system to be maintained or replaced without disturbing the permanently installed ORP probe.
Developed in collaboration with Prof. Dr. Tim Mansfeldt of the University of Cologne, the system is particularly suited to long-term studies where changes in soil oxidation and reduction conditions need to be monitored continuously.
Key Features
Continuous in-situ measurement of soil redox potential (Eh/ORP)
-1 to +1 V measurement range
1 mV resolution
Normalised 99.95% platinum measuring surface
Compact 6 mm diameter carbon-fibre shaft
Suitable for permanent soil installation
Suitable for saturated and unsaturated conditions
Separate Ag/AgCl reference electrode
Maintenance-free measuring electrode
Field and laboratory applications
Compatible with ecoTech enviLog monitoring systems
SDI-12 integration available via connection module
Can be combined with ecoTech soil pH electrodes in the same monitoring system
Applications
Ideal for soil science, wetland and peatland research, agricultural research, groundwater-influenced soils, lysimeter studies and long-term monitoring of changing oxidation and reduction conditions.
Specification | Details |
|---|---|
Measured parameter | Redox potential (EH / ORP) |
Measurement range | -1 to +1 V |
Resolution | 1 mV |
Measuring element | Platinum |
Platinum purity | 99.95% Pt, hard drawn |
Platinum element length | 5 mm |
Platinum element diameter | 1 mm |
Shaft material | Carbon fibre |
Shaft length | 300 mm |
Shaft diameter | 6 mm |
Standard cable length | 3 m |
Alternative lengths | Available on request |
Reference electrode | Separate reference electrode |
Installation | Permanent in-situ installation |
Suitable conditions | Unsaturated & water-saturated soils |
Application environment | Field & laboratory |
Data logging | Mobile readout or stationary monitoring |
Digital output* | SDI-12 via optional connection module |
pH integration | Can be combined with ecoTech soil pH electrodes |
Find Out More